OMEGA
HH507
Digital Thermometer
INTRODUCTION
This instrument is a 4½ digit, compact-sized portable digital thermometer
designed to use external K/J/T/E/R/S/N-type thermocouples as temperature
sensor.TemperatureindicationfollowsReferenceTemperature/VoltageTables
(N.I.S.T. Monograph 175 Revised to ITS-90) for K/J/T/E/R/S/N-type thermo-
couples. One K-type thermocouple are supplied with the thermometer.
SAFETY INFORMATION
Itisrecommendedthatyoureadthesafetyandoperationinstructionsbefore
using the thermometer.
WARNING
To avoid electrical shock, do not use this instrument when working
voltages at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature
measurement in microwave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple leads. To prolong
lead life, avoid sharp bends in the leads, especially near the connector.
The
symbol on the instrument indicates that the operator must refer to an
explanation in this manual.
1
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable
Measurement Range:
Thermocouple
K-TYPE
J-TYPE
Range
-200°C to 1372°C, -328°F to 2501°F
-210°C to 1050°C, -346°F to 1922°F
-200°C to 400°C, -328°F to 752°F
-210°C to 790°C, -346°F to 1454°F
0°C to 1767°C, 32°F to 3212°F
0°C to 1767°C, 32°F to 3212°F
-50°C to 1300°C, -58°F to 2372°F
T-TYPE
E-TYPE
R-TYPE
S-TYPE
N-TYPE
Resolution:
K-TYPE
0.1°C, 0.1°F (0°F to 700°F)
0.2°F on other range
0.1°C, 0.1°F (0°F to 500°F)
0.2°F on other range
0.1°C, 0.1°F (0°F to 600°F)
0.2°F on other range
0.1°C, 0.1°F (0°F to 450°F)
0.2°F on other range
1°C , 1°F
J-TYPE
T-TYPE
E-TYPE
R-TYPE
S-TYPE
N-TYPE
1°C , 1°F
0.1°C , 0.2°F
2
Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
K/J/T/E-TYPE
±(0.05% rdg + 0.3°C) -50°C to 1370°C
±(0.05% rdg + 0.7°C) -50°C to -210°C
±(0.05% rdg + 0.6°F) -58°F to 2498°F
±(0.05% rdg + 1.4°F) -58°F to -346°F
N-TYPE
±(0.05% rdg + 0.7°C) -50°C to 0°C
±(0.05% rdg + 0.3°C) 0°C to 1300°C
±(0.05% rdg + 1.4°F) -58°F to 32°F
±(0.05% rdg + 0.6°F) 32°F to 2372°F
R/S-TYPE
±(0.05% rdg + 2°C) 0°C to 1767°C
±(0.05% rdg + 4°F) 32°F to 3212°F
Temperature Coefficient: 0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Reading Rate: one time per second.
Input Connector: Accepts standard miniature thermocouple connectors (flat
blades spaced 7.9mm, center to center).
3
ENVIRONMENTAL
Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H.
Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display:4½digitliquidcrystaldisplay(LCD)withmaximumreadingof9999.9.
Overload: "----.-" is display.
Battery: Standard 9V battery.
Battery Life: 100 hours typical with carbon zinc battery.
Auto power off: The meter key switch inactive for more than 30 minutes,
press power key to resume operation.
Dimensions: 192mm(H) x 91mm(W) x 52.5mm(D).
Weight: 365g.
4
OPERATING INSTRUCTIONS
1
2
Power Switch
The key turns the thermometer on or off. In the MIN MAX record mode
can not power off, must leave MIN MAX record mode then power off.
°C/°F Selecting the Temperature Scale
Readings are dual displayed in either degrees Celsius(°C) or degrees
Fahrenheit(°F). When the thermometer is turned on, it is set to the
temperature scale that was in use when the thermometer was last turned
off. To change the temperature scale, press the °C/°F key.
3
HOLD Mode (only Main display)
Pressing the HOLD key to enter the Data Hold mode, the "HOLD" annun-
ciator is displayed. When HOLD mode is selected, the thermometer held
the present readings and stops all further measurements.
Pressing the HOLD key again to cancel HOLD mode causing thermometer
to resume taking measurements.
In the MIN/MAX recording mode, press HOLD key to stop the recording.
Press HOLD key again to resume recording. (Previously recorded reading
are not erased).
5
4 TIMER STOP-WATCH Mode
Press TIMER STOP-W key to enter stop-watch mode stop-watch is
displayedinseconddisplay.Pressstart/stopkeytotogglestop-watchstarts
and stops (Time goes up to 99 minutes & 59.9 seconds). In the this mode
the automatic power-off feature is disabled.
Pess Clear key to zero the stop-watch.
Press TIMER STOP-W key again to exit the stop-watch mode.
5
6
K/J/T/E/R/S/N T1 Input Thermocouple Type Select (Main
display)
The TYPE key circulating selects the K/J/T/E/R/S/N type thermocouple as
input.
When the thermometer is turned on, it is set to the type selected that was
in use when the thermometer was last turned off.
MIN MAX with Time record Mode (only Main display)
Press MIN MAX key to enter the MIN MAX Recording mode, (displays the
Maximum reading with time, Minimum reading with time and Average
reading stored in record mode). In the this mode the automatic power-off
feature is disabled and
key, °C/°F key, REL key, SET key, Hi/Lo Limits
key and TYPE key are disabled.
The beeper emits a tone when a new minimum or maximum value is
recorded.
6
Push MIN MAX key to cycle through the Max/Min/Avg/Present readings.
If an overload is recorded, the averaging function is stopped and average
value display"----.-".
The true average of all the reading taken over at least 22 hours preiod can
be displayed. If 22 hours is exceeded, new averages are no longer
calculated. Thelastcalcalatedvalueisretainedastheaveragereading, but
the actual minimum and maximum reading will continue to be captured.
In the this mode, press HOLD key to stop the recording of readings, all
values are frozen, press again to restart recording.
TopreventaccidentallossofMIN,MAXandAVGdata,inthismodecanonly
be cancelled by pressing and hold down the MIN MAX key for 2 seconds to
exit and erased recorded readings.
7
REL Relative mode (only Main display)
Pressing REL key to enter the Relative mode, zero the display, and store
the displayed Reading as a reference value and annunciator REL is
displayed. Press REL key again to exit the relative mode.
The relative value can also be entered by the user. (See "SET mode" later
in this manual.)
When the desired Relative value has been entered, press REL key to enter
the Relative mode, press SET key use set Relative value as a reference
value. Press REL key again to exit the relative mode.
In the Relative mode, the value (can not >±3000.0 counts) shown on the
LCD is always the difference between the stored reference and the present
reading.
7
8 Hi/Lo LIMITS mode (only Main display)
Press Hi/Lo LIMITS key to enter the Hi/Lo Limits comparative mode,
"LIMIT" is displayed. In the this mode the automatic power-off feature is
disabled and REL key, HOLD key, MAX/MIN key are disabled, when input
temperature value exceed Hi. The beeper emits a continuity pulse tone and
"Hi" is displayed, and when input temperature value exceed Lo value. The
beeper emits a discontinuous pulse tone and "Lo" is displayed. Press Hi/Lo
LIMIT key again to exit the Hi/Lo LIMIT mode.
9
SET mode (Relative value set, Time set and Hi/Lo Limits value
set and Time set)
1. Press SET key to enter Relative valuse SET mode (Press ENTER key can
escape relative valuse set mode), REL set mode. = = = =.= is displayed in
maindisplay.Relativevalueisenteredviaoverlaynumbers,whenyouwant
to get negative values push (- 0) key for end of numbers, then press overlay
ENTER key, stored the relative value, enter Hi/Lo Limits value set mode.
2. Hi Limit value set mode, (Press ENTER key can escape Hi Limit value set
mode), = = = =.= is displayed in main display, Hi Limit value is entered via
overlay numbers, when you want to get negative values push (- 0) key for
end of numbers, then press overlay ENTER key, stored the Hi Limit value,
enterLoLimitvaluesetmode(Press ENTERkeycanescapeLoLimitvalue
set mode), = = = =.= is displayed in main display, Lo Limit value is entered
via overlay numbers, when you want to get negative values push (- 0) key
for end of numbers, then press overlay ENTER key, stored the Lo Limit
value, enter Time set mode.
8
3. Time set mode, (Press ENTER key can escape Time set mode) = = : = =
= = is displayed in third display. Time (hours, minutes, seconds) value is
entered via overlay numbers, then press overlay ENTER key. Time start
from set time value and exit set mode.
10
Relative value Recall display mode
PressRELRECALLkeytodisplaytheRelativesetvalueonseconddisplay.
Press REL RECALL key again to exit this mode.
11
Hi/Lo Limits value Recall display mode
PressHi/LoRECALLkeytodisplaytheHisetLimitvalueonseconddisplay.
Press Hi/Lo RECALL key again to display the Lo Limits set value on second
display. Press Hi/Lo RECALL key again to exit this mode.
12 " " Button
Press " " button to toggle on and off of backlight. The backlight will switch-
off automatically after 60 seconds.
9
OPERATOR MAINTENANCE
WARNING
To avoid possible electrical shock, disconnect the thermocouple
connectors from the thermometer before removing the cover.
Battery Replacement
Powerissuppliedbya9volt"transistor"battery. The" "appearsontheLCD
display when replacement is needed. To replace the battery, remove the two
screws from the back of the meter and lift off the battery cover. Remove the
battery from battery contacts.
10
HH 506(R)/HH 507(R) CALIBRATION PROCEDURE
Note: The following calibration procedure should perform only by qualified technicians
who have access to the items as following.
Equipment: The class of calibrator had better 10 times greater than the measured meter.
1. Set the jumper position on J1 position.
2. Input DCV 17mV to the T1 after the display is stabilized (forward 4 digits), then
press “ HOLD “ key and "0.0" is displayed in main display.
3. Input DCV 60mV after the display is stabilized (first 4 digits), then press
“ HOLD “ key and "0.1" is displayed in main display.
4. Input 0LC (K-type) after the display is stabilized (first 4 digits), then press
“ HOLD “ key and "0.2" is displayed in main display.
5. Don't push any key and the meter will turn off automatically after 20 seconds.
6. After adjustment set the jumper back to J3 position.
7. Turn on the meter then input 0LC (K-type). The display reading reads 0°C if
the calibration procedure is right.
1
WARRANTY / DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months
from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product
warranty to cover handling and shipping time.This ensures that OMEGA's customers receive maximum coverage on each product.If
the unit should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Service Department will issue an
Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to
be defective it will be repaired or replaced at no charge. OMEGA's WARRANTY does not apply to defects resulting from any action
of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair,
or unauthorized modification.This WARRANTY isVOID if the unit shows evidence of having been tampered with or shows evidence
of being damaged as a result of excessive corrosion; or current, heat moisture or vibration; improper specification; misapplication;
misuse or other operating conditions outside of OMEGA's control. Components which wear are not warranted, including but not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products.However,OMEGA neither assumes responsibility
for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance
with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it
will be as specified and free of defects. OMEGA MAKES NO OTHERWARRANTIES OR REPRESENTATIONS OF ANY KIND
WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED WARRANTIES INCLUDING ANY
WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY:The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with
respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall
not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a "Basic Component" under 10
CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any
Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way,
OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and additionally, purchaser will
indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in
such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA'S
CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then
be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was PURCHASED.
FOR NON-WARRANTY REPAIRS, consult OMEGA for
current repair charges. Have the following information
available BEFORE contacting OMEGA:
2. Model and serial number of the product under warranty, and 1. P.O. number to cover the COST of the repair.
3. Repair instructions and/or specific problems relative to the
product.
2. Model and serial number of product , and
3. Repair instructions and/or specific problems relative to the
product.
OMEGA's policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest
in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 1999 OMEGA ENGINEERING,
INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or
machine-readable from, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.
Whe re Do I Find Eve rything I Ne e d for
Proc e ss Me a sure m e nt a nd Control?
OMEGA...Of Course !
TEMPERATURE
DATA ACQUISITION
þ Data Acquisition &
Engineering Software
þ Communications-Based
Acquisition Systems
þ Plug-in Cards for Apple, IBM
& Compatibles
þ Thermocouple, RTD & Thermistor
Probes, Connectors, Panels & Assemblies
þ Wire: Thermocouple, RTD & Thermistor
þ Calibrators & Ice Point References
þ Recorders, Controllers & Process Monitors
þ Infrared Pyrometers
þ Datalogging Systems
þ Recorders, Printers & Plotters
PRESSURE/ STRAIN AND FORCE
þ Transducers & Strain Gauges
þ Load Cells & Pressure Gauges
þ Displacement Transducers
HEATERS
þ Heating Cable
þ Instrumentation & Accessories
þ Cartridge & Strip Heaters
þ Immersion & Band Heaters
þ Flexible Heaters
FLOW/ LEVEL
þ Rotameters, Gas Mass Flowmeters
& Flow Computers
þ Laboratory Heaters
ENVIRONMENTAL
þ Air Velocity Indicators
þ Turbine/Paddlewheel Systems
þ Totalizers & Batch Controllers
MONITORING AND CONTROL
þ Metering & Control Instrumentation
þ Refractometers
p H/ CONDUCTIVITY
þ Pumps & Tubing
þ pH Electrodes, Testers & Accessories
þ Benchtop/Laboratory Meters
þ Controllers, Calibrators, Simulators
& Pumps
þ Air, Soil & Water Monitors
þ Industrial Water & Wastewater
Treatment
þ pH, Conductivity & Dissolved Oxygen
Instruments
þ Industrial pH & Conductivity Equipment
M-2813/0799
|